Poster · 2024 · Drug Discovery Chemistry

A Drug Discovery Platform Enabling Robust Intracellular Delivery of Cell Impermeable Cargo via Mechanoporation

Sophia Hirsch, Zhihui Song, Andrew Larocque, Rachel Conover, Athira Pillai, Jacquelyn Hanson, Alec Barclay, Mathias Pawlak, Armon Sharei

Cells
T cells, iPSCs, Cancer Cell Lines
Cargos
mRNA, CRISPR RNPs, DELs, Antibody, Oligo, Macrocycle, Peptide, small molecules

What this document says

This document describes Portal Biotechnologies' mechanoporation-based drug discovery platform and research-scale system for intracellular delivery of cell-impermeable cargo. The poster outlines a workflow in which cells are transiently disrupted by forcing them through defined pores at low pressure (3–15 PSI) using a MicroBooster™ consumable and Gateway™ benchtop instrument, enabling delivery of diverse cargos (small molecules, oligos, proteins, DEL probes, mRNA, RNPs) into a range of cell types. Data panels show improved delivery relative to electroporation for a JAK inhibitor, robust delivery of fluorescent DNA-encoded library probes into iPSCs and HeLa cells, co-delivery of antibody/oligo/dextran in HeLa, dose-responsive uptake of macrocycles, efficient delivery into HEK293 cells, compatibility with automated high-throughput dispensing, and multiplexed and sequential delivery (mRNA, dextran, CRISPR RNPs) in naive and activated T cells with maintained viability and morphology.

Introduction to Portal’s Approach: Implementing a Universal Cell Engineering Solution

Portal presents a platform intended to be compatible across many cell types and cargos, including small molecules, oligos, peptides, antibodies, DELs, mRNA, siRNA, DNA, RNPs, and proteins. The workflow uses low-pressure (3–15 PSI) mechanical deformation of cells through a membrane-format consumable (MicroBooster™) to transiently disrupt the plasma membrane and permit extracellular cargo to enter the cytosol before resealing. The system is embodied in a research-scale Gateway™ benchtop instrument and cell type–specific consumables, and is designed for integration with high-throughput robotics and early adopter use in 2024.

Background on Mechanoporation

Prior mechanoporation work shows that mechanical deformation of cells through narrow channels can enable cytosolic delivery of macromolecules to primary cells and cell lines. The general mechanism is that cells squeezed through a constriction experience temporary membrane disruption; cargo present extracellularly can enter the cytosol prior to membrane resealing. The poster cites earlier studies demonstrating this concept and positions Portal’s implementation as a low-pressure, membrane-format approach for high-throughput delivery.

Mechanoporation Can be Used to Deliver Membrane Impermeable Cargo to Cells

A cell-impermeable JAK pathway inhibitor identified in a cell-free screen was delivered into mixed PBMCs by either electroporation or mechanoporation. Flow cytometry comparing viability and compound delivery shows improved delivery using mechanoporation relative to electroporation, and functional activity was confirmed by knockdown of STAT in the mechanoporated cells, demonstrating intracellular access and downstream activity of the delivered small molecule.

Robust DNA Encoded Library Delivery in iPSCs and HeLa Cells

Mechanoporation was used to deliver a fluorescently tagged DNA-encoded library (DEL) probe into both iPSCs and HeLa cells. For each cell type, ‘Control’ (no boost) and ‘Boost’ (mechanoporated) conditions were compared by flow cytometry, with boosted samples showing shifted fluorescence histograms indicative of increased intracellular DEL probe signal immediately after boosting.

Efficient Co-delivery of Antibody, Oligos, and Dextran to HeLa Cells

HeLa cells were mechanoporated with a mixture containing a labeled isotype control antibody, a Cy5-labeled oligo, and fluorescent dextran. Bar charts for viability and delivery efficiency plus flow cytometry histograms show that boosted HeLa cells gained fluorescence for all three cargo types while maintaining high viability, indicating effective co-delivery of multiple cargo modalities.

Enabling High Throughput Screening by Mixing Cargo Post-Boost

HeLa cells were boosted without cargo and then added to wells pre-loaded with 3 kDa fluorescent dextran at varying times post-boost (5–40 seconds). Flow cytometry (N=3) shows that delivery is highest when cargo is added shortly after boosting and decreases with longer delays, consistent with a transient permeabilized state that reseals over time.

Enabling High Throughput Screening by Automated Dispensing into Pre-loaded Cargo Plates

An automated workflow integrating a booster cartridge with a single-head dispenser was used to deliver boosted HeLa cells into many wells of 96-well plates pre-loaded with fluorescent dextran. Flow cytometry traces from wells on multiple plates demonstrate consistent dextran delivery across plates. Phase-contrast images at 10 minutes and 24 hours post-boost show preserved cell morphology, supporting compatibility with automated high-throughput screening.

Macrocycle Delivery to HeLa cells Shows Robust Dose Response

HeLa cells were boosted and then plated into wells containing FITC-labeled macrocycle at concentrations up to 0.2 mg/mL. Flow cytometry measurements indicate a dose-dependent increase in intracellular fluorescence (delivery) while viability remains high, demonstrating control over delivered dose and a clear dose–response relationship for macrocycle uptake.

What the figures show

  • Schematic from cells/cargo to boosting to intracellular delivery: The image shows cells and extracellular cargo approaching a membrane-like constriction labeled 'Boosting.' Cells are depicted passing through the membrane at low pressure and becoming mechanically deformed; after passage, cargo is shown inside the cells. This schematic illustrates the core mechanism: transient membrane disruption from forcing cells through defined pores permits extracellular cargo to enter the cytosol before the membrane reseals.
  • Research-scale system and consumables illustration: Small drawings depict the Gateway™ benchtop instrument and the MicroBooster™ consumable cartridge (dimensions labeled 18 mm and 35 mm). Bullet text highlights that the Gateway™ controls delivery parameters and fits in a hood, while the MicroBooster™ has a defined pore size tailored to cell type. A note indicates low pressure allows integration into existing robotics for high-throughput use.
  • Efficient JAK compound delivery: A schematic and flow cytometry histograms compare electroporation (orange) and mechanoporation (blue) for delivering a cell-impermeable JAK inhibitor into mixed PBMCs. Histograms on the left show viability and compound delivery, and plots on the right show functional activity assessed by STAT knockdown. The blue mechanoporation traces are shifted relative to the orange electroporation traces, indicating improved delivery and confirmed downstream activity after mechanoporation.
  • DEL delivery in iPSCs and HeLa cells: For each cell type, paired panels labeled 'Control' and 'Boost' show flow cytometry histograms and images for a fluorescent DNA-encoded library probe. Boosted samples show a clear shift to higher fluorescence compared to controls, indicating enhanced intracellular DEL probe signal immediately after mechanoporation in both iPSCs and HeLa cells.
  • Co-delivery of antibody, oligo, and dextran to HeLa: Bar graphs present percent viability and percent delivery-positive cells for three cargo types (isotype antibody, Cy5-oligo, fluorescent dextran). Flow cytometry histograms show untreated (grey) versus boosted (colored) traces, with boosted samples exhibiting increased fluorescence for all cargos while viability bars remain high, demonstrating successful multiplex cargo delivery in HeLa cells.
  • Post-boost mixing time-course for HeLa cells: A timeline cartoon shows cells boosted with no cargo then pipetted into wells pre-loaded with 3 kDa Alexa Fluor 488 dextran at timepoints from 5 to 40 seconds. Flow cytometry histograms (N=3) show that delivery efficiency is highest at short delays after boost and declines as the delay increases, consistent with rapid membrane resealing.
  • Automated dispensing into pre-loaded plates: Illustration shows a booster cartridge integrated with a single-head dispenser adding boosted cells into 96-well plates pre-loaded with fluorescent dextran. Plots of flow cytometry data for wells from plate 2 and plate 7 show consistent delivery across plates. Phase contrast images at 10 minutes and 24 hours post-boost show preserved cell morphology, supporting compatibility with automated high-throughput workflows.
  • Macrocycle dose–response in HeLa cells: A schematic indicates cells boosted with no cargo then plated into wells containing FITC-labeled macrocycle at increasing concentrations (0–0.2 mg/mL). Bar graphs and fluorescence plots show dose-dependent increases in delivery efficiency while viability remains high, demonstrating controlled, dose-responsive intracellular uptake.
  • FITC-labeled macrocycle delivery to HEK293 cells: Bar charts show viability and delivery efficiency comparisons between untreated and boosted HEK293 cells. Fluorescence microscopy images (FITC and transmitted light) display bright FITC signal in boosted cells and little signal in untreated cells, indicating efficient macrocycle delivery with preserved cell health.
  • Multiplex delivery in naive T cells: A schematic and flow cytometry bar plots show naive T cells mechanoporated with GFP mRNA, fluorescent dextran, and B2M CRISPR RNPs. Flow cytometry 40 hours post-boost reports GFP expression and reduced B2M expression, illustrating simultaneous mRNA expression and CRISPR-mediated protein knockdown from multiplexed delivery.
  • Sequential delivery in T cells (unstimulated and activated): Panels depict protocols and flow cytometry outcomes for sequential boosting—first with TRAC RNPs, then after activation with multiplexed cargo (GFP mRNA and B2M RNPs). A plot indicates approximately 51.4% of cells achieved the combined phenotype (GFP expression plus knockouts) in an activated sequential experiment, demonstrating sequential mechanoporation for multiplex gene editing and expression.
  • Footer QR/code call-to-action: A boxed area contains a QR code and a call-to-action 'Connect with Portal! Contact us for beta program eligibility' and the URL www.portal.bio, inviting contact for participation in Portal's beta program.

The summary and sections were read from the page layout. They are not a word-for-word extract, and they include only what was visible.