Anil Narasimha, Sophia Hirsch, Darby Kreienberg, Andrew Larocque, Eleni Rogers, Zhihui Song, Xi Ai, James Vasta, Matthew Robers, Alec Barclay, Armon Sharei
- Cells
- T cells, iPSCs, PBMCs, Cancer Cell Lines
- Cargos
- mRNA, CRISPR RNPs, circRNA, small molecules, Polymers
What this document says
This document presents Portal Biotechnologies' silicon membrane-based mechanoporation platform and instruments (Gateway™, MicroBooster™ cartridge, and Galaxy™) for mechanically delivering membrane‑impermeable cargos into live cells. It provides examples across cell types (HeLa, HEK293, iPSCs, PBMCs, activated and unstimulated T cells, MiaPaCa-2) showing delivery of small molecules, fluorescent probes, DNA-encoded library (DEL) probes, macrocycles, mRNA, circRNA, proteins (LgBiT), and CRISPR RNPs. Assays demonstrate improved delivery versus electroporation for a JAK‑STAT inhibitor, reproducible dispensing into pre-loaded multiwell plates including inline robotic workflows, dose responses for macrocycle uptake, intracellular target engagement and BRET assays with impermeable tracers, intracellular protein detection via LgBiT complementation, and functional engineering of T cells (CD19 CAR and mbIL-12) and multiplexed delivery in unstimulated T cells. The poster emphasizes compatibility with research and high-throughput workflows and highlights preserved cell viability and assay readouts following mechanoporation.
Introduction to Portal’s Approach: Implementing a Universal Cell Engineering Solution
Portal describes a silicon membrane-based mechanoporation workflow intended to be compatible across many cell types and cargo classes (small molecules, oligos, peptides, antibodies, DELs, mRNA, siRNA, DNA, RNPs, proteins). Cells are boosted through a membrane at low operating pressure (3–15 PSI) using a bench-top Gateway™ unit or high-throughput Galaxy™ system and MicroBooster™ consumable cartridges with defined pore sizes. The systems are shown as compatible with liquid handlers and in-line robotics to enable research- and screening-scale delivery.
Background on Mechanoporation
The poster summarizes prior mechanoporation concepts where mechanical deformation through narrow channels transiently disrupts the cell membrane, permitting cytosolic delivery of macromolecules to primary cells and cell lines before membrane resealing. References to earlier literature are provided to contextualize mechanical delivery as an established approach for intracellular access.
Mechanoporation Can be Used to Deliver Membrane Impermeable Cargo to Cells
An experiment delivered a cell‑impermeable JAK‑STAT inhibitor (identified in a cell‑free screen) into mixed PBMCs by either electroporation or mechanoporation. Flow cytometry assessed viability and compound delivery, and activity was confirmed by STAT knockdown. The mechanoporation condition showed improved delivery relative to electroporation and resulted in STAT knockdown, supporting intracellular activity of the delivered inhibitor.
Enabling High Throughput Screening by Dispensing into Pre-loaded Multiwell Cargo Plates
FITC‑labeled oligo and Cascade blue‑labeled 3 kDa dextran were dispensed into 96‑well plates. HeLa cells were mechanoporated in a trough on the Galaxy system and then dispensed across five 96‑well plates with a multichannel. Flow cytometry from plates 1 and 5 shows consistent delivery across plate columns, indicating reproducible delivery across multiple plates in this workflow.
Robust DNA Encoded Library Delivery in iPSCs and HeLa Cells
iPSCs and HeLa cells were mechanoporated with a fluorescently tagged DNA encoded library (DEL) probe and analyzed immediately by flow cytometry. Boosted (mechanoporated) conditions in both cell types show increased fluorescent signal versus controls, indicating effective DEL probe delivery by mechanoporation.
Consistent Delivery of Cargo Across Pre-loaded Plates by Automated In line dispensing
A cartridge designed for in-line integration with a robotic single‑head dispenser enabled automated dispensing of mechanoporated HeLa cells into sequential pre-loaded 96‑well plates. Flow cytometry for fluorescent dextran delivery across Plate 2 and Plate 7 demonstrates consistent delivery across wells and later plates in the run. Phase contrast images at 10 minutes and 24 hours post‑boost show cell morphology, supporting cell health after boosting.
Macrocycle Delivery to HeLa cells Shows Robust Dose Response
HeLa cells were mechanoporated and dispensed into wells pre-loaded with FITC‑labeled macrocycle across a concentration range (0–0.2 mg/mL / 0–96 µM). Flow cytometry demonstrates a dose‑dependent increase in macrocycle uptake in boosted cells while untreated cells show little to no fluorescence; viability metrics indicate boosted cells remain viable across doses.
Efficient Delivery of Labeled Macrocycle to HEK293 Cells
HEK293 cells mechanoporated with a FITC‑labeled macrocycle were assessed by flow cytometry and fluorescence microscopy. Boosted cells show strong FITC signal and maintained viability compared to untreated cells, and microscopy confirms intracellular localization of the labeled macrocycle in boosted samples.
What the figures show
- Efficient JAK Compound Delivery / Corresponding STAT Knockdown: Two schematic plots compare electroporation (orange) and mechanoporation (blue) for delivering a cell‑impermeable JAK‑STAT inhibitor into mixed PBMCs. The left overlay indicates improved delivery with mechanoporation versus electroporation as measured by flow cytometry, and the right plot shows STAT knockdown across conditions. The figure demonstrates that mechanoporation enabled greater intracellular delivery of the inhibitor and that the delivered compound produced STAT knockdown, supporting intracellular activity after boosting.
- Plate delivery to HeLa cells: A line chart plots percent delivery to HeLa cells across 96‑well plate columns for Plate 1 and Plate 5 after mechanoporation in a trough and dispensing into pre‑loaded wells containing fluorescent cargos (FITC oligo and Cascade blue dextran). The plotted series for both plates are consistent across columns, indicating reproducible delivery across wells and plates. This illustrates the Galaxy system's ability to mechanoporate cells and dispense them into multiple pre‑loaded plates for high‑throughput screening.
- DNA Encoded Library Delivery in iPSCs and HeLa Cells: Two panels labeled 'iPSC Boost' and 'HeLa Boost' show fluorescence distributions or percent positive cells comparing control versus boosted conditions. Boosted samples in both cell types have increased fluorescence relative to controls, indicating immediate uptake of a fluorescent DEL probe via mechanoporation and demonstrating delivery in both stem cells and a cancer cell line.
- In-line automated dispensing workflow schematic: A small schematic depicts a mechanoporation cartridge integrated with a robotic dispenser that sequentially dispenses boosted cells into pre‑loaded 96‑well plates. This diagram demonstrates how mechanoporation can be coupled to automation for high‑throughput workflows, enabling boosted cells to be delivered into many plates with robotic handling.
- Dextran Delivery to HeLa Cells by Automated Dispensing: Line plots show percent dextran delivery across wells for Plate 2 and Plate 7 at 10 minutes and 24 hours post‑boost. Delivery percentages remain consistent across wells and between early and later plates in the run, indicating stable performance during an automated sequence. Accompanying phase contrast images at 10 min and 24 h (scale bar = 300 μm) show cell morphology, supporting maintained cell health after boosting.
- Macrocycle Dose Response in HeLa Cells: Bar charts and flow cytometry histograms display FITC‑macrocycle uptake across increasing concentrations (0–96 μM) and compare boosted versus untreated cells. Boosted cells show a dose‑dependent increase in fluorescence while untreated cells show little uptake; viability bars indicate boosted cells remain viable across doses. This demonstrates mechanoporation enables quantitative, dose‑responsive intracellular delivery of macrocycles.
- Macrocycle Delivery to HEK293 Cells (flow and microscopy): A bar graph compares percent macrocycle‑positive cells and viability for untreated vs boosted HEK293 cells, and paired fluorescence microscopy images (transmitted light and FITC) show intracellular green fluorescence in boosted cells. Boosted samples exhibit strong intracellular FITC signal and preserved viability, visually confirming delivery and cell health.
- Intracellular LgBiT Detection of KRAS(G12C) Degradation: A schematic and bar graph illustrate MiaPaCa‑2 cells with endogenous HiBiT‑tagged KRAS(G12C) mechanoporated with LgBiT and treated with PROTAC LC‑2 (0–10 μM). Luminescence measured after Nano‑Glo Endurazine addition decreases with higher LC‑2 concentrations, consistent with target degradation reducing HiBiT/LgBiT complementation signal. This shows mechanoporation can deliver protein reagents for intracellular complementation assays detecting degradation kinetics.
- BRET Assays with Impermeable Kinase Tracer (concentration response): A dose‑response curve shows BRET signal increasing with tracer concentration in HeLa cells expressing SRC‑NanoLuc after mechanoporation of the impermeable tracer, with an inset schematic of NanoLuc‑SRC and tracer producing BRET. The curve indicates intracellular tracer binding to SRC yields measurable BRET and demonstrates tracer concentration dependence.
- BRET Competition with Kinase Inhibitors: Bar plots/curves compare BRET signals when cells are pre‑incubated with Ponatinib or Bosutinib versus control, showing reduced BRET in inhibitor‑treated samples. This indicates inhibitors compete with the impermeable tracer for SRC binding inside cells, demonstrating the assay's ability to measure intracellular target engagement and competition.
- Mechanoporated vs Digitonin mRNA BRET Profiles: Two grouped bar charts compare BRET profiles following NanoLuc mRNA delivery by mechanoporation versus digitonin permeabilization. The profiles are similar between the two methods, suggesting mechanoporation yields comparable assay readouts for mRNA delivery relative to chemical permeabilization.
- Engineered Activated T cells: Flow cytometry plots show expression of CD19 CAR and membrane‑bound IL‑12 in activated T cells 48 hours after mechanoporation with corresponding circRNAs. A cytotoxicity plot displays killing of Raji target cells at varying effector:target ratios, with engineered T cells showing increased cytotoxicity. These figures demonstrate simultaneous delivery of multiple circRNA constructs and functional CAR activity post‑boost.
The summary and sections were read from the page layout. They are not a word-for-word extract, and they include only what was visible.
# Enabling novel live cell assays with a silicon membrane-based platform for mechanical delivery of impermeable small molecules, peptides, and probes to accelerate early drug discovery
This page is a text version of a Portal Biotechnologies publication. The bibliographic record is labeled in the text so it can be read without the surrounding website.
## Bibliographic record
- Title: Enabling novel live cell assays with a silicon membrane-based platform for mechanical delivery of impermeable small molecules, peptides, and probes to accelerate early drug discovery
- Document type: Poster
- Year: 2025
- Outlet: SLAS (Europe)
- Cells: T cells, iPSCs, PBMCs, Cancer Cell Lines
- Cargos: mRNA, CRISPR RNPs, circRNA, small molecules, Polymers
- Page URL: https://portal.bio/publications/2025-enabling-novel-live-cell-assays-with-a-silicon-membrane-based-platform-for-mechanical--ajkF-hEAACkAc9zQ
- Markdown URL: https://portal.bio/publications/2025-enabling-novel-live-cell-assays-with-a-silicon-membrane-based-platform-for-mechanical--ajkF-hEAACkAc9zQ.md
- Original file: https://portal-bio.cdn.prismic.io/portal-bio/aFmSDnfc4bHWioTD_SLASEuPoster.pdf
## Portal description
Anil Narasimha, Sophia Hirsch, Darby Kreienberg, Andrew Larocque, Eleni Rogers, Zhihui Song, Xi Ai, James Vasta, Matthew Robers, Alec Barclay, Armon Sharei
## Summary
This document presents Portal Biotechnologies' silicon membrane-based mechanoporation platform and instruments (Gateway™, MicroBooster™ cartridge, and Galaxy™) for mechanically delivering membrane‑impermeable cargos into live cells. It provides examples across cell types (HeLa, HEK293, iPSCs, PBMCs, activated and unstimulated T cells, MiaPaCa-2) showing delivery of small molecules, fluorescent probes, DNA-encoded library (DEL) probes, macrocycles, mRNA, circRNA, proteins (LgBiT), and CRISPR RNPs. Assays demonstrate improved delivery versus electroporation for a JAK‑STAT inhibitor, reproducible dispensing into pre-loaded multiwell plates including inline robotic workflows, dose responses for macrocycle uptake, intracellular target engagement and BRET assays with impermeable tracers, intracellular protein detection via LgBiT complementation, and functional engineering of T cells (CD19 CAR and mbIL-12) and multiplexed delivery in unstimulated T cells. The poster emphasizes compatibility with research and high-throughput workflows and highlights preserved cell viability and assay readouts following mechanoporation.
## Sections
### Introduction to Portal’s Approach: Implementing a Universal Cell Engineering Solution
Portal describes a silicon membrane-based mechanoporation workflow intended to be compatible across many cell types and cargo classes (small molecules, oligos, peptides, antibodies, DELs, mRNA, siRNA, DNA, RNPs, proteins). Cells are boosted through a membrane at low operating pressure (3–15 PSI) using a bench-top Gateway™ unit or high-throughput Galaxy™ system and MicroBooster™ consumable cartridges with defined pore sizes. The systems are shown as compatible with liquid handlers and in-line robotics to enable research- and screening-scale delivery.
### Background on Mechanoporation
The poster summarizes prior mechanoporation concepts where mechanical deformation through narrow channels transiently disrupts the cell membrane, permitting cytosolic delivery of macromolecules to primary cells and cell lines before membrane resealing. References to earlier literature are provided to contextualize mechanical delivery as an established approach for intracellular access.
### Mechanoporation Can be Used to Deliver Membrane Impermeable Cargo to Cells
An experiment delivered a cell‑impermeable JAK‑STAT inhibitor (identified in a cell‑free screen) into mixed PBMCs by either electroporation or mechanoporation. Flow cytometry assessed viability and compound delivery, and activity was confirmed by STAT knockdown. The mechanoporation condition showed improved delivery relative to electroporation and resulted in STAT knockdown, supporting intracellular activity of the delivered inhibitor.
### Enabling High Throughput Screening by Dispensing into Pre-loaded Multiwell Cargo Plates
FITC‑labeled oligo and Cascade blue‑labeled 3 kDa dextran were dispensed into 96‑well plates. HeLa cells were mechanoporated in a trough on the Galaxy system and then dispensed across five 96‑well plates with a multichannel. Flow cytometry from plates 1 and 5 shows consistent delivery across plate columns, indicating reproducible delivery across multiple plates in this workflow.
### Robust DNA Encoded Library Delivery in iPSCs and HeLa Cells
iPSCs and HeLa cells were mechanoporated with a fluorescently tagged DNA encoded library (DEL) probe and analyzed immediately by flow cytometry. Boosted (mechanoporated) conditions in both cell types show increased fluorescent signal versus controls, indicating effective DEL probe delivery by mechanoporation.
### Consistent Delivery of Cargo Across Pre-loaded Plates by Automated In line dispensing
A cartridge designed for in-line integration with a robotic single‑head dispenser enabled automated dispensing of mechanoporated HeLa cells into sequential pre-loaded 96‑well plates. Flow cytometry for fluorescent dextran delivery across Plate 2 and Plate 7 demonstrates consistent delivery across wells and later plates in the run. Phase contrast images at 10 minutes and 24 hours post‑boost show cell morphology, supporting cell health after boosting.
### Macrocycle Delivery to HeLa cells Shows Robust Dose Response
HeLa cells were mechanoporated and dispensed into wells pre-loaded with FITC‑labeled macrocycle across a concentration range (0–0.2 mg/mL / 0–96 µM). Flow cytometry demonstrates a dose‑dependent increase in macrocycle uptake in boosted cells while untreated cells show little to no fluorescence; viability metrics indicate boosted cells remain viable across doses.
### Efficient Delivery of Labeled Macrocycle to HEK293 Cells
HEK293 cells mechanoporated with a FITC‑labeled macrocycle were assessed by flow cytometry and fluorescence microscopy. Boosted cells show strong FITC signal and maintained viability compared to untreated cells, and microscopy confirms intracellular localization of the labeled macrocycle in boosted samples.
## What the figures show
- Efficient JAK Compound Delivery / Corresponding STAT Knockdown: Two schematic plots compare electroporation (orange) and mechanoporation (blue) for delivering a cell‑impermeable JAK‑STAT inhibitor into mixed PBMCs. The left overlay indicates improved delivery with mechanoporation versus electroporation as measured by flow cytometry, and the right plot shows STAT knockdown across conditions. The figure demonstrates that mechanoporation enabled greater intracellular delivery of the inhibitor and that the delivered compound produced STAT knockdown, supporting intracellular activity after boosting.
- Plate delivery to HeLa cells: A line chart plots percent delivery to HeLa cells across 96‑well plate columns for Plate 1 and Plate 5 after mechanoporation in a trough and dispensing into pre‑loaded wells containing fluorescent cargos (FITC oligo and Cascade blue dextran). The plotted series for both plates are consistent across columns, indicating reproducible delivery across wells and plates. This illustrates the Galaxy system's ability to mechanoporate cells and dispense them into multiple pre‑loaded plates for high‑throughput screening.
- DNA Encoded Library Delivery in iPSCs and HeLa Cells: Two panels labeled 'iPSC Boost' and 'HeLa Boost' show fluorescence distributions or percent positive cells comparing control versus boosted conditions. Boosted samples in both cell types have increased fluorescence relative to controls, indicating immediate uptake of a fluorescent DEL probe via mechanoporation and demonstrating delivery in both stem cells and a cancer cell line.
- In-line automated dispensing workflow schematic: A small schematic depicts a mechanoporation cartridge integrated with a robotic dispenser that sequentially dispenses boosted cells into pre‑loaded 96‑well plates. This diagram demonstrates how mechanoporation can be coupled to automation for high‑throughput workflows, enabling boosted cells to be delivered into many plates with robotic handling.
- Dextran Delivery to HeLa Cells by Automated Dispensing: Line plots show percent dextran delivery across wells for Plate 2 and Plate 7 at 10 minutes and 24 hours post‑boost. Delivery percentages remain consistent across wells and between early and later plates in the run, indicating stable performance during an automated sequence. Accompanying phase contrast images at 10 min and 24 h (scale bar = 300 μm) show cell morphology, supporting maintained cell health after boosting.
- Macrocycle Dose Response in HeLa Cells: Bar charts and flow cytometry histograms display FITC‑macrocycle uptake across increasing concentrations (0–96 μM) and compare boosted versus untreated cells. Boosted cells show a dose‑dependent increase in fluorescence while untreated cells show little uptake; viability bars indicate boosted cells remain viable across doses. This demonstrates mechanoporation enables quantitative, dose‑responsive intracellular delivery of macrocycles.
- Macrocycle Delivery to HEK293 Cells (flow and microscopy): A bar graph compares percent macrocycle‑positive cells and viability for untreated vs boosted HEK293 cells, and paired fluorescence microscopy images (transmitted light and FITC) show intracellular green fluorescence in boosted cells. Boosted samples exhibit strong intracellular FITC signal and preserved viability, visually confirming delivery and cell health.
- Intracellular LgBiT Detection of KRAS(G12C) Degradation: A schematic and bar graph illustrate MiaPaCa‑2 cells with endogenous HiBiT‑tagged KRAS(G12C) mechanoporated with LgBiT and treated with PROTAC LC‑2 (0–10 μM). Luminescence measured after Nano‑Glo Endurazine addition decreases with higher LC‑2 concentrations, consistent with target degradation reducing HiBiT/LgBiT complementation signal. This shows mechanoporation can deliver protein reagents for intracellular complementation assays detecting degradation kinetics.
- BRET Assays with Impermeable Kinase Tracer (concentration response): A dose‑response curve shows BRET signal increasing with tracer concentration in HeLa cells expressing SRC‑NanoLuc after mechanoporation of the impermeable tracer, with an inset schematic of NanoLuc‑SRC and tracer producing BRET. The curve indicates intracellular tracer binding to SRC yields measurable BRET and demonstrates tracer concentration dependence.
- BRET Competition with Kinase Inhibitors: Bar plots/curves compare BRET signals when cells are pre‑incubated with Ponatinib or Bosutinib versus control, showing reduced BRET in inhibitor‑treated samples. This indicates inhibitors compete with the impermeable tracer for SRC binding inside cells, demonstrating the assay's ability to measure intracellular target engagement and competition.
- Mechanoporated vs Digitonin mRNA BRET Profiles: Two grouped bar charts compare BRET profiles following NanoLuc mRNA delivery by mechanoporation versus digitonin permeabilization. The profiles are similar between the two methods, suggesting mechanoporation yields comparable assay readouts for mRNA delivery relative to chemical permeabilization.
- Engineered Activated T cells: Flow cytometry plots show expression of CD19 CAR and membrane‑bound IL‑12 in activated T cells 48 hours after mechanoporation with corresponding circRNAs. A cytotoxicity plot displays killing of Raji target cells at varying effector:target ratios, with engineered T cells showing increased cytotoxicity. These figures demonstrate simultaneous delivery of multiple circRNA constructs and functional CAR activity post‑boost.
## Layout reading
Title (top, large):
Enabling novel live cell assays with a silicon membrane-based platform for mechanical delivery of impermeable small molecules, peptides, and probes to accelerate early drug discovery
Authors and affiliations (below title):
Anil Narasimha1, Sophia Hirsch1, Darby Kreienberg1, Andrew Larocque1, Eleni Rogers1, Zhihui Song1, Xi Ai2, James Vasta3, Matthew Robers3, Alec Barclay1, Armon Sharei1
1 Portal Biotechnologies, 134 Coolidge Ave, Watertown, MA 02472, USA. www.portal.bio
2 Merck Research Laboratories, 2015 Galloping Hill Rd, Kennilworth, NJ, 07033
3 Promega Corporation, 2800 Woods Hollow Rd, Madison, WI 53711 USA
Section: Introduction to Portal’s Approach: Implementing a Universal Cell Engineering Solution
Subheaders and short labeled items (left-to-right across the top area):
- Cells and Cargo
- Mechanical Disruption via Boosting
- Intracellular Delivery
- Research-Scale System
- Gateway™ (image of a compact bench-top unit)
- MicroBooster™ delivery cartridge (schematic showing cell type-specific consumable for optimal mechanoporation, defined pore size for deformation)
- High-Throughput Capabilities
- Galaxy™ (image of a larger instrument)
- “Powered by Portal” Consumables (schematic of consumables)
Bulleted text under this section:
- Portal is creating a transformative technology for cellular engineering compatible across cell types and cargos
- Small molecules, oligos, peptides, antibodies, DELs, mRNA, siRNA, DNA, RNPs, proteins, and other cargos
- Primary and patient-derived cell lines including immune and stem cells
- Cells are boosted through a membrane at low operating pressure (3-15 PSI)
- Membrane format1 enables high-throughput cellular engineering
- Compact bench-top unit that controls delivery parameters
- Galaxy system for use with liquid handlers
- Compatible with systems like the Opentrons Flex
- In-line integrations with existing robotics
- Compatible with systems like the LGR PDII
Reference footnote:
1. Williams AR, Bao S, Miller DL. Biotechnol Bioeng. 1999
Section: Background on Mechanoporation (left column)
Bulleted points:
- Previous work has shown that mechanical deformation of cells through a microfluidic channel enables cytosolic intracellular delivery1. Delivers macromolecules2,3,4 to both primary cells and cell lines
- Cells are squeezed through a narrow channel at high pressure. The cell membrane is temporarily disrupted, allowing the target cargo to enter the cytosol before the membrane seals5.
References listed:
1. Sharei, et al., PNAS, 2013
2. Szeto, et al., Scientific Reports, 2015
3. Sharei, et al., PLOS One, 2015
4. Lee and Sharei, et al. Nano Letters 2012
5. Sharei et al., Integrative Biology 2014
Section heading (right of Background):
Mechanoporation Can be Used to Deliver Membrane Impermeable Cargo to Cells
Subfigure label: Efficient JAK Compound Delivery (small header)
Subfigure label: Corresponding STAT Knockdown (small header)
Text block describing the experiment:
- A cell impermeable JAK-STAT pathway inhibitor was identified in a cell-free screen. This small molecule was delivered to mixed PBMCs using either electroporation (orange) or mechanoporation (blue).
- Flow cytometry was used to assess viability and delivery of the JAK compound into the cells (left).
- Activity of the inhibitor was confirmed by assessing knockdown of STAT in the mixed PBMCs (right).
Caption credit:
Li J, et al. 2017. ACS Chemical Biology.
Figure (visual description and plain-language explanation):
- What is drawn: Two schematic-like plots are presented. The left shows a mixed PBMC population with an overlay indicating "improved delivery using mechanoporation" comparing electroporation (orange) and mechanoporation (blue). The right shows a line/area style plot of STAT knockdown across conditions.
- What is compared: Delivery and activity of a cell-impermeable JAK-STAT inhibitor delivered by electroporation (orange) versus mechanoporation (blue) into mixed PBMCs.
- What the result shows: The figure indicates that mechanoporation improved delivery (blue greater than orange) and that STAT knockdown was observed, supporting activity of the delivered inhibitor.
- How the depicted process works: The small molecule identified in cell-free screens was delivered intracellularly either by electroporation or by mechanically disrupting cells via mechanoporation, then flow cytometry assessed delivery and downstream STAT knockdown.
Section: Enabling High Throughput Screening by Dispensing into Pre-loaded Multiwell Cargo Plates (center-left)
Subheaders:
- Independent Instrument
- Instrument and Consumable
Experiment description (text):
- FITC-labeled oligo, and Cascade blue-labeled 3 kDa dextran were mixed and dispensed into 96-well plates. The Galaxy system was used to mechanoporate HeLa cells into a trough before dispensing into wells of 5 96-well plates with a multichannel. Flow cytometry was conducted post-boost to assess delivery of the cargos to the cells. Data from plates 1 and 5 shown here.
Figure: Plate delivery to HeLa cells (chart)
Plain-language paragraph describing chart:
- What is drawn: A line chart plotting percent delivery to HeLa cells on the y-axis across plate columns on the x-axis. Two data series correspond to Plate 1 and Plate 5; small icons of 96-well plates are shown.
- What is compared: Cargo delivery efficiency across wells in pre-loaded 96-well plates (specifically Plate 1 vs Plate 5).
- What the result shows: The plotted points indicate delivery was consistent across the plate columns for both Plate 1 and Plate 5, suggesting reproducible delivery across multiple plates.
- How the depicted process works: HeLa cells are mechanoporated in a trough, then dispensed into pre-loaded wells containing fluorescent cargos; flow cytometry is used to quantify percent of cells receiving fluorescent cargo from each well.
Section: Robust DNA Encoded Library Delivery in iPSCs and HeLa Cells (upper right)
Subheaders and small panels:
- iPSC Boost / HeLa Boost
Left text:
- Induced pluripotent stem cells (iPSCs) or HeLa cells were mechanoporated with a fluorescently-tagged DNA encoded library (DEL) probe. Delivery efficiency was measured using flow cytometry to detect the fluorescent tag immediately after boosting.
Figures (bar/violin/histogram like panels):
Plain-language paragraph describing visuals:
- What is drawn: Two sets of plots comparing control versus boost conditions for iPSCs and HeLa cells, showing fluorescence distributions or percent positive cells. One panel labeled "iPSC Boost" and another "HeLa Boost."
- What is compared: Delivery of fluorescent DEL probe in mechanoporated (Boost) versus untreated control cells for both iPSCs and HeLa.
- What the result shows: In both cell types, the Boost condition shows increased fluorescent signal relative to control, indicating improved delivery of the DEL probe by mechanoporation.
- How the depicted process works: Cells were mechanoporated with the fluorescent DEL probe, then immediate flow cytometry quantified uptake.
Section: Consistent Delivery of Cargo Across Pre-loaded Plates by Automated In line dispensing (middle left)
Subheaders and small schematic:
- In-line Integrations
- Consumable only
Experiment description text:
- A cartridge was designed for in-line integration of mechanoporation with a robotic single-head dispenser. HeLa cells were automatically dispensed into 96-well plates pre-loaded with fluorescent dextran. Flow cytometry was used to assess dextran delivery. Data for delivery across the 2nd (wells 97-192) and 7th (wells 577-672) 96-well plates are shown here. Phase contrast images show morphology of HeLa cells 10 minutes and 24 hours post-boost. Scale bar (white) = 300 μm.
Figures and plain-language paragraphs:
1) Schematic of workflow (small image):
- What is drawn: A schematic showing a mechanoporation cartridge integrated with a robotic dispenser that dispenses boosted cells into pre-loaded 96-well plates in sequence.
- What is compared: Not a comparison; demonstrates an automated inline dispensing workflow.
- What the result shows: Illustrates the capability to integrate boosting with automated dispensing into multiple plates for high-throughput assays.
- How the depicted process works: Cells are mechanoporated in the cartridge, then a robot dispenses cells into pre-loaded plates in series.
2) Graph: Dextran Delivery to HeLa Cells by Automated Dispensing (line chart):
- What is drawn: Line plots showing percent dextran delivery across wells for Plate 2 and Plate 7 at 10 min post-boost and 24 h post-boost; plotted points per well column.
- What is compared: Delivery performance between Plate 2 and Plate 7 and over time (10 min vs 24 h).
- What the result shows: Delivery percentages are consistent across wells and across later plates (Plate 7), indicating stable performance through a run; images show cells appear healthy at 10 min and 24 h.
- How the depicted process works: Boosted HeLa cells are dispensed into wells pre-loaded with dextran; flow cytometry assesses delivery at stated timepoints. The phase contrast images (300 μm scale bar) show cell morphology shortly after boost and after 24 h.
Section: Macrocycle Delivery to HeLa cells Shows Robust Dose Response (center)
Text and schematic:
- HeLa cells were boosted then manually added into wells of a 96-well plate pre-loaded with FITC-labeled macrocycle at 0-0.2 mg/ml. Flow cytometry was conducted post-boost to assess viability and efficiency of macrocycle delivery into boosted cells.
Figures and plain-language descriptions:
1) Dose response bar chart (multiple bars labeled 0 µM, 12 µM, 24 µM, 48 µM, 96 µM):
- What is drawn: Bar charts showing percent delivery or fluorescence intensity for HeLa cells at increasing macrocycle concentrations, plus flow cytometry histogram panels comparing untreated vs Boost for specific concentration(s).
- What is compared: Macrocycle delivery across a dose range (0 to 96 µM) and between boosted and untreated cells.
- What the result shows: Boosted cells show increased uptake of the FITC-labeled macrocycle in a dose-dependent manner; untreated cells show little to no fluorescence. Viability bars indicate boosted cells remain viable across doses (exact numeric values are in the plotted bars).
- How the depicted process works: HeLa cells are mechanoporated and then exposed to wells with increasing concentrations of FITC-labeled macrocycle; delivery and viability are measured by flow cytometry.
Section: Efficient Delivery of Labeled Macrocycle to HEK293 Cells (right)
Experiment summary:
- HEK293 cells were mechanoporated with a FITC-labeled macrocycle. Viability and macrocycle delivery to the untreated and boosted samples were measured by flow cytometry (left) and fluorescence microscopy (right). Scale bar = 300 μm.
Figures and plain-language paragraph:
- What is drawn: A bar graph comparing percent macrocycle-positive cells and viability for untreated vs Boosted HEK293 cells, and fluorescence microscopy images (transmitted light and FITC) showing cells with green fluorescence in boosted condition.
- What is compared: Untreated vs Boosted HEK293 cells for macrocycle uptake and viability.
- What the result shows: Boosted HEK293 cells show strong FITC signal indicating macrocycle delivery while maintaining viability; microscopy visually confirms intracellular FITC-labeled macrocycle in boosted cells, not in untreated.
- How the depicted process works: Mechanoporation enables the FITC-labeled macrocycle to enter HEK293 cells; flow cytometry quantifies delivery and viability, microscopy shows localization.
Section: Intracellular Protein Detection with Promega’s LgBiT Reagents (lower left)
Small schematic/flow:
- Left icon sequence labeled: Endogenous protein with HiBiT -> LgBiT delivery for luminescence -> Degradation removes signal
Experimental text:
- MiaPaCa-2 cells endogenously tagged with HiBiT at the KRAS(G12C) locus were treated with the PROTAC LC-2 for 24 hours (0-10 μM) and mechanoporated with LgBiT protein (Promega)
- Nano-Glo® Endurazine™ was added and luminescence was detected on a GloMAX plate reader to detect degradation of KRAS(G12C)
- Potential utility in other intracellular labeling assays to determine PPIs, degradation and other live cell kinetics
Figure(s) and plain-language paragraph:
- What is drawn: A bar graph plotting luminescence (relative units) against LC-2 concentration (0-10 μM) showing decreasing luminescence with increasing LC-2 in the mechanoporated samples.
- What is compared: Luminescence signal corresponding to KRAS( G12C ) HiBiT tagging with LgBiT complementation across LC-2 doses.
- What the result shows: Increasing concentrations of the PROTAC LC-2 reduce luminescence signal, consistent with degradation of KRAS(G12C) detected by intracellular LgBiT complementation introduced by mechanoporation.
- How the depicted process works: Cells with HiBiT-tagged endogenous protein are mechanoporated with LgBiT; when present, LgBiT complements HiBiT to produce luminescence; PROTAC-mediated degradation reduces the HiBiT-tagged target and thus luminescence.
Section: Intracellular Target Engagement Assays with Impermeable Molecules (center-lower)
Top panel text:
- (Top) HeLa cells were transfected with SRC-NanoLuc Fusion vector then mechanoporated with varying amounts of impermeable kinase tracer. NanoGlo substrate was added and BRET (Bioluminescence resonance energy transfer) was detected on a plate reader.
Bottom panel text:
- (Bottom) HeLa cells transfected with SRC-NanoLuc Fusion vector were incubated for 2 hours with Src kinase inhibitors (Ponatinib or Bosutinib) then mechanoporated with an impermeable kinase tracer. NanoGlo substrate was added and BRET was detected on a plate reader.
Figures and plain-language paragraphs:
1) Top BRET concentration-response plot:
- What is drawn: A dose-response curve plotting BRET signal versus concentration of kinase tracer, with an inset schematic showing the NanoLuc-SRC fusion and tracer producing BRET upon binding.
- What is compared: BRET signal intensity across tracer concentrations in mechanoporated cells.
- What the result shows: BRET signal increases with tracer concentration, indicating tracer engagement with SRC inside cells after mechanoporation.
- How the depicted process works: Mechanoporation delivers a normally impermeable fluorescent kinase tracer into cells expressing a NanoLuc-tagged target; upon tracer binding, energy transfer (BRET) to NanoLuc substrate produces measurable signal.
2) Bottom competitive inhibitor plot:
- What is drawn: Bar plots or curves showing reduced BRET when cells are pre-treated with kinase inhibitors Ponatinib or Bosutinib compared to control.
- What is compared: BRET signal in presence vs absence of competitive inhibitors.
- What the result shows: Pre-incubation with inhibitors reduces BRET from the impermeable tracer, indicating target engagement by cell-permeable inhibitors and demonstrating the assay can be used for intracellular target engagement.
- How the depicted process works: Inhibitors compete with the impermeable tracer for SRC binding inside cells; when inhibitor binds, less tracer binds and BRET signal decreases.
Section: Mechanoporated mRNA Yields Similar BRET Profile Compared to Digitonin-permeabilized Control (lower-right)
Text:
- NanoLuc mRNA was introduced by either mechanoporation (top) or permeabilization with digitonin (bottom) and BRET signal was measured using a plate reader.
Figure and plain-language paragraph:
- What is drawn: Two bar charts or grouped bar plots comparing BRET profiles for mechanoporated versus digitonin-permeabilized cells across conditions; a small schematic indicates mechanoporation versus digitonin.
- What is compared: BRET signal resulting from NanoLuc mRNA introduction by mechanoporation compared to digitonin permeabilization.
- What the result shows: BRET profiles are similar between mechanoporation and digitonin conditions, indicating mechanoporation yields comparable intracellular expression/assay readouts to chemical permeabilization while likely preserving cell integrity better.
- How the depicted process works: mRNA encoding NanoLuc is delivered inside cells either by mechanical membrane disruption (mechanoporation) or by chemically permeabilizing the membrane (digitonin); after expression or immediate substrate addition, BRET is measured.
Section: Activated T cells Engineered with Simultaneous circRNA Delivery for IL-12 and CD19 CAR (lower-left)
Text and figure labels:
- Human Pan-T cells were isolated and activated for 6 days. Activated T cells were mechanoporated with CD19 CAR circRNA and membrane bound IL-12 (mbIL-12) circRNA. Boosted cells were co-cultured with Raji cells at different ratios of effector cells (E) to target cells (T). Flow cytometry was conducted 48h post-boost to assess expression of CD19 CAR and mbIL-12 (left) and cytotoxicity (right).
Figures and plain-language paragraphs:
1) Flow cytometry plots for CD19 CAR and mbIL-12 expression:
- What is drawn: Scatter/contour plots showing populations gated for expression markers; bar or dot plots quantifying percent positive cells for CD19 CAR and mbIL-12 for control vs boosted samples.
- What is compared: Expression of delivered circRNA constructs (CD19 CAR and mbIL-12) in activated T cells after mechanoporation compared to control.
- What the result shows: Boosted T cells show expression of CD19 CAR and mbIL-12, indicating successful simultaneous delivery of multiple circRNA constructs by mechanoporation.
- How the depicted process works: Activated T cells are mechanoporated with circRNA encoding desired constructs; expression is measured by flow cytometry after 48 hours.
2) Cytotoxicity assay (killing of Raji target cells):
- What is drawn: A dose-response or bar chart showing cytotoxicity of engineered T cells at different effector:target ratios (E:T), with boosted cells showing higher cytotoxicity compared to controls.
- What is compared: Cytotoxic activity of control vs engineered T cells across E:T ratios.
- What the result shows: Mechanoporated T cells expressing CD19 CAR exhibit increased killing of Raji target cells, demonstrating functional activity of delivered constructs.
- How the depicted process works: Engineered T cells are co-cultured with target cells and target cell viability or lysis is measured to assess CAR T cell function.
Section: Robust Multiplex Delivery in Unstim. T Cells (center-lower)
Text:
- Unstimulated T cells isolated from PBMCs were mechanoporated with multiplexed GFP mRNA, fluorescently-labeled dextran polymers, and β2 microglobulin (B2M) CRISPR RNPs. Flow cytometry was performed after 40hr.
Figure and plain-language paragraph:
- What is drawn: Flow cytometry dot plots showing GFP expression vs dextran uptake and B2M knockout (B2M expression) for control vs boosted cells; bar graphs quantify percent positive/knockout.
- What is compared: Multiplexed delivery efficiency of three cargo types in unstimulated T cells: GFP mRNA, fluorescent dextran, and CRISPR RNP targeting B2M.
- What the result shows: Boosted unstimulated T cells show simultaneous delivery: GFP expression, dextran fluorescence, and B2M editing (reduced B2M expression) indicating robust multiplexed intracellular delivery.
- How the depicted process works: Mechanoporation transiently disrupts membranes allowing multiple cargo types to enter the cytosol; outcomes measured by flow cytometry after 40 hours.
Section: Summary (bottom-right)
Bullet points:
- Portal’s mechanoporation technology can be used for drug discovery and cell engineering
- Compatible across broad cell types including: PBMCs, T cells, B cells, NK cells, HSCs, iPSCs
- Can be used to deliver single or multiplexed cargoes including: DELs, Peptides, Macrocycle, Antibodies, Small molecules
- Portal’s Research Scale Gateway™ and High Throughput Galaxy™ systems enable complex cell engineering and delivery of cell impermeable molecules with a simple workflow
- Small molecules, peptides, and proteins can be delivered into cells at scale for use in downstream screening
- Contact information: Connect with Portal! Contact us for beta program eligibility; www.portal.bio
Additional visual elements and images (general description and plain-language paragraphs):
- Top-center and right show product images: Gateway™ (compact bench-top), MicroBooster™ cartridge (diagram showing consumable and pore sizes), and Galaxy™ instrument (larger high-throughput system). These images depict the hardware used to perform mechanoporation at research and high-throughput scales.
- What is drawn: Photorealistic renderings or photos of three instruments and a schematic of the cartridge/consumable.
- What is compared: Size and intended application (gateway for bench scale, galaxy for high-throughput).
- What the result shows: The company offers multiple scales of instrumentation and consumables designed for mechanoporation workflows.
- How the depicted process works: Cells and cargo are combined, cells pass through a silicon membrane in the cartridge under low pressure (3-15 PSI), membrane mechanically disrupts cells briefly allowing cargo entry, membrane reseals.
- Several small microscopy images across the poster (phase contrast and FITC fluorescence) showing cells 10 min and 24 h post-boost, and FITC-labeled macrocycle in HEK293 cells.
- What is drawn: Brightfield/phase contrast images and matched fluorescence images showing cellular morphology and intracellular fluorescence.
- What is compared: Untreated vs boosted cells and early vs later time points.
- What the result shows: Boosted cells display intracellular fluorescence when given labeled cargo and maintain morphology over 24 h; untreated cells show little or no fluorescence.
- How the depicted process works: Mechanoporation allows fluorescent cargo to enter cells; microscopy confirms intracellular localization and cell health post-treatment.
Notes about accuracy and limits:
- All transcribed headings, labels, figure descriptions, and experimental summaries have been taken verbatim from the poster text where quoted and otherwise described plainly from the figures as printed.
- No claims beyond what is visible on the poster have been added.
## Source note
The summary and sections were read from the page layout. They are not a word-for-word extract, and they include only what was visible.